Go to the documentation of this file.
1#pragma once
2
3#include <cstdint>
4
5namespace fsw::calibration {
6
7/**
8 * A 4-20 mA PT's wire code -> the shunt code the conversions and calibrations work in.
9 *
10 * The code is full scale at 2^31 against the 2.5 V internal reference, and the loop at
11 * 20 mA across 120 Ohm is 2.4 V, so every reading a live loop can produce sits in
12 * [0, 2^31). A code at or above 2^31 is the ADC's two's-complement NEGATIVE: the shunt
13 * below 0 V, which is what an open or unpowered loop's noise does around zero.
14 *
15 * It used to be clamped to 2^31-1 instead -- full scale, 20 mA, 5000 psi. On 2026-11-07
16 * COPV-PT-R and HPD-PT (PT board 2, connectors 1 and 3, both reading ~0 mA) produced
17 * codes like 0xFFFF0FB5 (-61515) a few times a minute; each became a 5000 psi sample,
18 * the output EMA turned it into a 1250 psi pulse decaying over ~1 s, and the controller
19 * and the plots both received it.
20 *
21 * A negative code reads as no current: 0, which the conversion reports as 0 psi.
22 */
23inline uint32_t hp_pt_shunt_code(uint32_t wire) {
24 return wire > 0x7FFFFFFFu ? 0u : wire;
25}
26
27} // namespace fsw::calibration
Output smoothing for calibrated PT pressure, per sensor uid (ESP32 and LabJack T7 alike).
uint32_t hp_pt_shunt_code(uint32_t wire)
A 4-20 mA PT's wire code -> the shunt code the conversions and calibrations work in.
Definition HpPtCode.hpp:23